What Is Time-Domain Reflectometry?_
Time-domain reflectometry relates an electrical reflection to its return time from a defined reference plane. For a segment with approximately uniform propagation velocity, if delta t is the outward-and-return time, one-way distance L = vp x delta t / 2 = c x VF x delta t / 2. VF is the dimensionless velocity factor relative to the speed of light c; an NVP stated as a percentage must first be represented as that fraction. Dividing time by velocity does not give a distance.
Technical Details
Original fictional TDR-120 stipulates a 120 ns round trip from the relevant reference plane and VF = 2/3. Using the deliberately rounded c = 0.300 m/ns gives vp = 0.200 m/ns and L = 0.200 x 120 / 2 = 12 m. Keeping the same time but assuming VF = 0.80 gives 14.4 m, a 20% change in the calculated location without an observed cable change. These are invented inputs and arithmetic, not cable specifications, instrument readings or a repair location. If a separate record already reports a one-way distance of 12 m, do not halve that distance again. Confirm the axis definition, reference plane and actual cable-specific NVP before interpreting a real record. NVP uncertainty, time resolution, the measurement arrangement and instrument suitability remain relevant; a calibration record alone does not validate those inputs or identify the physical cause of a reflection. No connection, test setup or fault-finding procedure is provided.
Sources
- Tektronix: TDR Test
Checked 2026-10-26. Undated web primer; round-trip delay and resolution concepts only. No setup, diagnosis, operating procedure or supported installation inferred.
- Fluke Networks: Length measurements for the DSX CableAnalyzer Series Testers
Checked 2026-10-26. May 11, 2015 product note: returned delay, cable-specific NVP and its uncertainty only. Historical standards allowances, default values and software instructions are not adopted.
How Leviathan Systems Works with Time-Domain Reflectometry
Leviathan can discuss agreed physical observations, cable identification, labeling checks, physical-connectivity checks, defect records and handoff documentation. Formal instrument work requires confirmed suitable equipment, qualified personnel, method, coverage, acceptance criteria and required records for the specific engagement. No every-link certification, instrument inventory, manufacturer endorsement, electrical or liquid-side testing, or full network/software commissioning is implied.
Related service
Physical-Layer Verification →